Skip to main content
Fig. 2 | Military Medical Research

Fig. 2

From: Wave-shaped microfluidic chip assisted point-of-care testing for accurate and rapid diagnosis of infections

Fig. 2

Dimensions of channels in WMC components. a WMC microchannels longitudinal part is investigated with the designed width of 400 μm and characterized, and the actual width is 395.37 μm with a dimensional deviation of 1.16%. b Curved channel of WMC is also measured according to the width design of 400 μm. But the actual width is 405.41 μm, and the dimensional deviation is 1.35%. c Depth of the channel in WMC has a designed height of 400 μm. But in reality, the height is 407.25 μm with a dimensional deviation of 1.81%. d Compared with the designed 400 μm width of channels in the TM valve, it has 402.73 μm and 403.35 μm. Here the mean dimensional deviation is 0.61%. WMC wave-shaped microfluidic chip, TM valve translate-type mechanical valve